研磨人类牙齿 - 用SPEX冷冻研磨机研磨-文献综述-资讯-生物在线

研磨人类牙齿 - 用SPEX冷冻研磨机研磨

作者:德可納利(上海)仪器有限公司 2009-04-06T00:00 (访问量:6002)

用SPEX冷凍研磨機研磨人类牙齿.pdf用SPEX冷凍研磨機研磨人类牙齿
JAAS www.rsc.org/jaas
Dario Santos Junior, Fernando Barbosa Junior, Samuel Simiao de Souza and Francisco Jose Krug
Cryogenic sample grinding for copper, lead and manganese
determination in human teeth by slurry sampling GFAAS
用SPEX冷冻研磨机研磨人类牙齿而後用悬浮液进样石墨炉原子吸收法測試铜、 铅、锰

A simple method is proposed for copper, lead and manganese determination in deciduous teeth by graphite furnace atomic absorption spectrometry (GFAAS) using slurry sampling introduction and cryogenic sample preparation. Teeth samples were ground in a cryogenic mill in two steps: pre-cooling (5 min) and cryogenic grinding (2 min) in liquid nitrogen. After grinding, 90% of the sample particles were lower than 150 mm. The
minimum mass necessary for slurry preparation as an indicator of sub-sample homogeneity was evaluated by weighting masses between 5 and 20 mg directly in autosampler cups, followed by addition of 1 mL of a solution containing 0.04% Triton1 X-100 and 0.2% v/v HNO3. Samples (20 mL) were sonicated during 20 s, before delivering into a W–Rh coated platform. Detection limits based on integrated absorbance were 34.0 ng g21 Pb, 7.4 ng g21 Mn and 18.0 ng g21 Cu for 2% m/v slurries. W–Rh permanent modifier permitted calibration against aqueous standards. The Certified Reference Material (H-5 animal bone) from the International Agency of Atomic Energy (IAEA) was analyzed to determine lead for method validation. For copper, lead and manganese, 12 human teeth samples were analyzed using the proposed method with
calibration against aqueous solution and using a comparative Pd/Mg method with the same digested samples, and standard addition calibration, with no statistical difference at 95% level on applying the t-test.



德可納利(上海)仪器有限公司 商家主页

地 址: 徐汇区襄阳南路500号2509室

联系人: 李

电 话: 021-51690600

传 真: 021-51079676

Email:info@mpt-usa.com;sales@mpt-usa.com

相关咨询
ADVERTISEMENT